USER
MANUAL
38
51335-006
EEG System
Appendix 6: Manufacturer’s Declaration
EMC Compatibility
This section contains specific information regarding the device’s compliance with IEC 60601-1-2 and EN 60601-1-2.
The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the equipment as replacement parts for internal
components, may result in increased emissions or decreased immunity of the equipment.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided here.
The equipment or system should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the R40, including cables
specified by Lifelines Ltd. Otherwise, degradation of the performance of this equipment could result.
When in close proximity to the R40 Amplifier, do not use mobile phones, transmitters, power
transformers, motors, or other equipment that generates magnetic fields.
Accessory name
Type
Length
Description
USB Interface Cable
USB
2.8 m
USB shielded cable
Input electrodes
EEG disc electrodes
1 m
Unshielded EEG disc
electrodes
Input electrodes (E-cap)
EEG disc electrodes
1 m
Unshielded EEG disc
electrodes
Nonin XPOD
Shielded
2 m
Nonin
Aux. Connector cable
Shielded
1 m
Shielded cable
Patient Event Switch
CM-5
2 m
Two-core unshielded
cable
Appendix 6
Summary of Contents for neuro R40