USER
MANUAL
8
51335-006
EEG System
The conductive part of connectors and transducers should not contact other conductive parts
including earth. Always ensure that the transducer fitted is suitable for a connection of type BF
Isolation.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the Appendix.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the R40 EEG System,
including cables specified by Lifelines Ltd. Otherwise, degradation of the performance of this
equipment could result.
When in close proximity to the R40 EEG System, do not use mobile phones, transmitters, power
transformers, motors, or other equipment that generates magnetic fields. Refer to the Appendix for
more information.
Before transportation, the system should be put into its transportation position to reduce the risk of
overbalancing: disconnect mains supply cable; lower work top to lowest position; fully retract arms;
release brakes.
Do not exceed the Safe Additional Working Load of 4 kg on the worktop.
Only use approved sensors as specified by Lifelines.
The Amplifier must only be used with the USB cable provided with the unit.
Do not allow any liquid to enter the case of the instrument or connector. Do not use acetone on any
of the instruments.
Federal (USA) law restricts this device to sale by or on the order of a physician.
CONTRAINDICATIONS: There are no known contraindications to the use of this equipment.
Overview and Technical Description