The entire R40 system including the PC can be used within the patient environment. The mains leakage currents
and safety and regulatory requirements are met by the use of the medical power supply connected to the PC. The
current recommended PC is the Lenovo ThinkCentre Tiny All-In-One system.
The PC
1. Fit the Lenovo Tiny PC into the Lenovo monitor in accordance with the installation instructions provided with
the PC.
2. Fit the monitor to the VESA mount. Route the DC Power cable from the power supply, up inside the VESA mount
to the monitor. Connect the DC power cable to the power input of the All-in-one monitor.
3. Connect the mouse and keyboard to the rear of the PC in accordance with the installation instructions provided
with the PC. These cables can be neatly routed through the VESA mount pole.
4. Connect the orange power cord to the medical-grade power supply.
5. Connect the country specific power cord to orange mains power cord.
The laptop must only be connected to the medical-grade laptop power supply supplied or
authorised by Lifelines. Do not use a standard laptop power supply.
Only use the laptop supplied or authorised by Lifelines.
The Cart, Arm and Stand
This allows all the system components to be conveniently located on a custom cart. This cart has a height-
adjustable worktop, large lockable wheels, a medirail, VESA display mount and an accessory tray.
NOTE:
The cart is supplied flat-packed - refer to the enclosed instructions for assembly details.
A single mains input cable to the cart is provided, fitted with an earthed 3-pin IEC plug. A country-specific power
cord is provided for different territories. The connector on the cart’s mains cable is a locking type.
The cart should be positioned in use so that disconnection of the mains plug from the supply socket
can be accomplished without difficulty.
The arm will mount on the cart and accommodates either the Photic or the R40 on its bracket. The arm is also
supplied with a desk-clamp which allows it to be mounted on any convenient flat surface.
When the R40 EEG system is in operation, do not push the cart. Place the cart in the transportation
position before moving it.
Before transportation, the system should be put into its transportation position to reduce the risk of
overbalancing:
y
Switch off and disconnect mains cable.
y
Lower work top to lowest position.
y
Fully retract arms.
y
Release brakes.
y
Push the cart using the handle on the worktop.