USER
MANUAL
13
51335-006
EEG System
Oximeter Sensor
An optional oximeter sensor attaches to the patient’s finger
Patient Event pushbutton
The Patient Event Pushbutton is used by the patient to mark the instance of a significant event.
Aux DC Inputs
The Auxiliary DC inputs are intended for the connection of patient-attached transducers which are passive or
battery-powered such as sleep sensors. They must be insulated with no accessible conductive parts.
USB Cable
The Amplifier plugs directly into a USB port on the PC.
The Amplifier must only be used with the USB cable provided with the unit.
Medical grade AC/DC mains power supply module for PC
Where EEG studies are conducted within the patient environment the leakage current must be controlled. The
PC power supply supplied by Lifelines is a special medical-grade type (with appropriate safety standards), which
limits the mains leakage current of the system. It is rated for 90V – 230V input voltage.
Only use the laptop and the medical-grade power supply as supplied or authorised by Lifelines.
Do not use the standard laptop power supply
The Setup and Recording Software
The R40/Trackit setup software runs under Microsoft Windows XP, Windows Vista, Windows 7, 8 or 10 on the host
PC and is used to setup and review the R40 Amplifier and to record on to the PC.
Functions of the software:
y
Setup the input signals.
y
Setup and download the recording template.
This includes which electrodes are used and the recording montage.
y
Perform a calibration check of the Amplifier.
y
Perform an Impedance check on the Amplifier.
y
Perform an EEG recording.
y
View on-going EEG traces.
y
Review an EEG Recording.
Refer to the Trackit Plus software manual for more details.
Overview and Technical Description