![LiCONiC StoreX Series Operating Manual Download Page 90](http://html1.mh-extra.com/html/liconic/storex-series/storex-series_operating-manual_1907204090.webp)
StoreX
Series
STX_V078.doc
89
CMa,
23.11.2009
11. Maintenance
11.1 Maintenance
Guidelines
Handler 4 can only be used in a clean and dust free environment (laboratory
use). Dust can damage the open rotating parts. If the instrument is stored in a
dusty, unclean atmosphere for a long period of time, the dust must be removed
and the instrument must be cleaned very carefully before the instrument is
installed. Caution! There will be parts in need of new grease.
The Handler 4 is constructed for minimal maintenance and with the exception of
the periodic control of belt tension, there is no need for regular service.
The lift assembly requires regular but minimal check-up and maintenance.
Scheduling of maintenance is dependant on the frequency of movement.
Maintenance involves inspection of the lubrication of the gears,
mark of oil-
filling
in the lift assembly and of play in the gears.
Check
Maint.
Replace
Mo.
Mov.
Mo.
Mov.
Mo.
Mov.
Lift assembly
•
Lubrication film over the
entire length of the lift
rails
•
Free of particles
•
Must not indicate wear
•
Remove old lubrication
film
•
Apply new lubrication
film
•
Play of the vertical lift
guidance means parts
need to be changed
2 20000
6 50000
-- --
Shovel Guidance
•
Check max. movement
in height-play and side-
play
•
Lubrication film over
the entire length of the
linear-guidance
•
Free of particles
•
Remove old lubrication
film
•
Apply new lubrication
film
•
Too much play in the
linear guidance means
parts need to be
changed
2 20000
6 50000
-- --
Turn Drive Belt
•
Check for wear
•
Check belt teeth for
cracks
•
Check belt metal core
for free metal parts
2 20000
6 50000
12 --
Climate Chamber
Cleaning
•
Condenser Cleaning
•