![LiCONiC StoreX Series Operating Manual Download Page 11](http://html1.mh-extra.com/html/liconic/storex-series/storex-series_operating-manual_1907204011.webp)
StoreX
Series
STX_V078.doc
10
CMa,
23.11.2009
2.4
LiCONiC StoreX Climate Overview
Climate
IC Incubator
~33..50°C
>95%
RH
HC
Humid Cooler
04..20°C
90..95% RH
DC
Dry
Cooler
04..20°C
<5%RH
HR
Humid
Range
04..70°C
90..95%RH
DR
Dry
Range
04..50°C
<5%RH
DF
Deep
Freezer
-20°C
The graphical chart below depicts an overview of the various climate ranges
covered by each climate option. The vertical axis shows the relative humidity
while the horizontal axis shows temperature. It can be clearly seen that the units
cover both extremes: extreme humidity and extreme dryness.
HR-units are Wide Range instruments that cover a wide temperature range. They
are a combination, with integrated heater and cooler. Note that all cooled StoreX
have an integrated active cooling system. No additional external cooler is
needed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com