StoreX
Series
STX_V078.doc
37
CMa,
23.11.2009
Check the shovel height. The lower surface of the shovel should be
approximately 1-2 mm above the top surface of the Transfer Plate. The shovel
should be high enough to allow movement without touching the Transfer Plate
and low enough to pass under a plate placed on the Transfer Plate. Use screw
(5) to optimize the Transfer Plate height. By selecting option ‘S’ (Capital-S) you
can move the shovel back and forth continuously. This will allow you to observe
the shovel adjustment more precisely.
After the mechanical adjustments there are software related constants that
determine the location of the Transfer Station.
DM22 Data Memory 22 Defines the height of the lower level of the Transfer
Station for the lift axis
DM26 Data Memory 26 Defines the travel path for lifting or lowering a plate
at the Transfer Station
DM82 Data Memory 82 Defines the radial position at the Transfer Station for
the run axis
These constants can only be changed by special software or by overwriting them
on every cold start before initialization. Please refer to ‘Remote Operation’ for
details.
The STXFER program will let you read the values for these constants. Under
‘Actual Turn Position‘ in the main screen of the STXFER program you may read
the position of the Turn motion, while under ‘Actual Lift Position’ the position of
the lift motion is displayed. The value for DM22 is given at the lower transfer
position. The value for DM26 is calculated by subtracting the lower transfer
position from the upper transfer position.
6.2 Electrical
Connections
The electrical connections are at the rear of the instrument. Power and
communication connectors are on the same panel. The Transfer Station
connector feeds and controls the Transfer Station Option. The StoreX unit
automatically detects the presence of a motorized Transfer Station through this
connector. The Gas Connector connects to an optional gas supply such as CO2,
Oxygen or Nitrogen. Beware of the maximum gas pressure of 6 bar.
CAUTION !
Switch the System off before connecting and disconnecting the LiCONiC
StoreX
Disconnect the RS-232 interface only with the power switched off
Limit maximum length of communication cable to ~10 meters
Use current converters or amplifiers for longer distances or when operating
in a high field emissions environment
Transfer Station
Connector
Handling RS
232 Port
Option Port
Gas Inlet
Power Switch
Power
Connector
Fuse Holder
Voltage
Selector