![LiCONiC StoreX Series Operating Manual Download Page 20](http://html1.mh-extra.com/html/liconic/storex-series/storex-series_operating-manual_1907204020.webp)
StoreX
Series
STX_V078.doc
19
CMa,
23.11.2009
Response Time
~30 s
Warm-Up Time
5 min. operation
MTBF 30’000
hrs
3.5.4
O2 Oxygen Option
The sensors for oxygen measurement are based on the same principles as
described above. The main difference is the measurement range of the sensor
used in the O2 option.
Specifications
Measuring Range
0.1..95 % Vol.
Input Pressure
2
8
bar typ.
bar max.
Gas Flow
~10 l/min max.
Gas Consumption @ 25% O2, 20°C
<__
~__
~__
l/h standby
l/h 2-min.-access cycles
l/h 30sec.-access cycles
Accuracy ±0.5
%
FS
Stability
±0.5 % FS over 12 month
Repeatability
±1 % at full scale
Response Time
~30 s
Warm-Up Time
5 min. operation
MTBF 30’000
hrs
WARNING !
O2-Option leads to increased explosion risk
The maximum concentration of oxygen allowed with this option is limited to 25%
O2 vol. for safety reasons.
3.5.5
Other Gassing Options
There are other gassing options available. Please contact Liconic applications for
further information.
3.6
UHS (Ultra High Speed) Option
Applications with very short access times. The access speed of the UHS handler
is almost double compared to standard High Speed handlers. Access time for
UHS STX40 units is maximum ~12 seconds and for larger units access time is
maximum ~20 seconds. Please ask for the latest specifications for your model.
3.7 Customized
Color
Standard coating of the StoreX is RAL 9016. Other customized coating is
possible. Please specify RAL code and/or send sample. Note that this option will
increase delivery time.