![LiCONiC StoreX Series Operating Manual Download Page 55](http://html1.mh-extra.com/html/liconic/storex-series/storex-series_operating-manual_1907204055.webp)
StoreX
Series
STX_V078.doc
54
CMa,
23.11.2009
The option of remote access to the gas concentration allows a simple tracing of
the gas concentration over long periods of time.
The safety features are summarized in a table given below.
Feature Comment
F
1
O
2
C
3
Pressure Limiter
Limits internal pressure to some extent
+ + +
Gas Flow Limiter
Limits maximum flow avoiding full gas flow
+ + +
Exhaust
Channels exhaust gas
+ + -
4
Gas Switch
Turns gas flow on and off
-
+ +
Door Monitoring
Cuts gas flow when door opens
-
+ +
Gate Monitoring
Cuts gas flow when Gate is open over a
certain period of time
- + +
Gas Flow Timeout
Cuts gas when a continuous gas flow occurs
over a long period of time
- + +
Gas Flow Lookup
Controls gas flow according to system status
-
+ +
Gas Control
Controls gas flow according to measured gas
concentration values
- - +
Gas Measuring
Gas sensor measures actual gas
concentration
- - +
Gas Display
Gas concentration is displayed to user
-
- +
Gas Transmission
Remote gas concentration readout
-
- +
WARNING!
Because of the high safety standards maintained within the StoreX the
highest risk of failure or unwanted gas emission lies in theexternal
installations
1
Continuous Flow System
2
Open Loop System
3
Closed Loop System
4
Available on request
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com