Section 3
3-32
Setup and Operation
Total System Delay Examples
Total System Delay (ms) = Delay Time + (Delay Step
×
Delay Step Increment)
Software
Release
DAC (3.0.0)
SDM, RS-232
(3.0.0)
Delay Time (ms)
240
186
Delay Step
Increment
9
17
Total Delay (ms)
299
297
Delay Step (ms)
6.5
6.5
Bandwidth
Bandwidth (5, 10 or 20 Hz) determines the signal averaging done by the digital filter. To
avoid aliasing (only a concern for co-spectra, not for fluxes), one should sample the LI-7500
at a frequency greater than or equal to 2 times the bandwidth. Thus, if you are sampling at 10
Hz, set Bandwidth to 5 Hz.
NOTE: Bandwidth applies to all LI-7500 outputs (SDM, DACs, and RS-232).
Bandwidth is the frequency at which the indicated amplitude is 0.707 of the real amplitude
(Figure 3-4).
Summary of Contents for LI-7500
Page 1: ...Analyzer Environmental Division Instruction Manual 7500...
Page 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Page 41: ...Section 3 3 12 Setup and Operation...
Page 43: ...Section 3 3 14 Setup and Operation...
Page 46: ...Section 3 Setup and Operation 3 17...
Page 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Page 56: ...Section 3 Setup and Operation 3 27...
Page 58: ...Section 3 Setup and Operation 3 29...
Page 67: ...Section 3 3 38 Setup and Operation...
Page 70: ...Section 3 Setup and Operation 3 41...
Page 107: ...Section 6 6 6 Troubleshooting...