Appendix D
D-28
Appendix D
997e-3)(Z 6.7999997e-3)(A 4.5109792e3)(B 2.9099099e6)(C 8.9501600e7))(Band
(A 1.
1499999))))
(Outputs ?)lf causes a Outputs record to be put in the output queue.
Example Response:
(Outputs (BW 10)(Delay 0)(SDM (Address 7))(Dac1 (Source NONE)(Zero -5e-
2)(Full 4
e-1))(Dac2 (Source PRESSURE)(Zero -1e-1)(Full 4e-1))(RS232 (Baud
38400)(Freq 0)(
Pres TRUE)(Temp TRUE)(Aux TRUE)(Cooler TRUE)(CO2Raw TRUE)(CO2D TRUE)(H2ORaw
TRUE
)(H2OD TRUE)(Ndx TRUE)(DiagVal TRUE)(DiagRec FALSE)(Labels FALSE)(EOL
"0D0A")))
(Data ?)lf causes a Data record to be put in the output queue. (Note that an ENQ (0x05) will
do the same thing...)
Example Response:
(Data (Ndx 2471)(DiagVal 250)(CO2Raw 1.6319131e-1)(CO2D 3.5119712e1)(H2ORaw
3.16
72954e-2)(H2OD 1.7067077e2)(Temp 2.3874512e1)(Pres 9.8735609e1)(Aux
0)(Cooler 1.
5630015))
(Diagnostics ?)lf causes a Diagnostics record to be put in the output queue.
Summary of Contents for LI-7500
Page 1: ...Analyzer Environmental Division Instruction Manual 7500...
Page 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Page 41: ...Section 3 3 12 Setup and Operation...
Page 43: ...Section 3 3 14 Setup and Operation...
Page 46: ...Section 3 Setup and Operation 3 17...
Page 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Page 56: ...Section 3 Setup and Operation 3 27...
Page 58: ...Section 3 Setup and Operation 3 29...
Page 67: ...Section 3 3 38 Setup and Operation...
Page 70: ...Section 3 Setup and Operation 3 41...
Page 107: ...Section 6 6 6 Troubleshooting...