Section 3
Setup and Operation
3-23
Y
Slope = A
B
0
V
Figure 3-3. Response curve for external pressure or temperature sensors.
Use the temperature and pressure radio buttons to select the method for measuring these
values. If you are using the auxiliary input, enter the values for the multipler (A) and offset
(B). If you have updated user-entered values, you need to send them to the LI-7500 for
implementation using the
Update Selected
button. Click on
Cancel All
to restore any edited
values in this page to their original state.
Using Auxiliary Inputs
The LI-7500 has two auxiliary input channels that can be used to attach either an external
temperature or pressure sensor. Only one active external sensor can be attached and active at
a time. The LI7500 software (Inputs Tab) assumes selection of either the internal on-board
sensors, selection of one externally measured sensor in combination with one on-board sensor
or manually entered values of temperature and or pressure with one of these other options.
Summary of Contents for LI-7500
Page 1: ...Analyzer Environmental Division Instruction Manual 7500...
Page 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Page 41: ...Section 3 3 12 Setup and Operation...
Page 43: ...Section 3 3 14 Setup and Operation...
Page 46: ...Section 3 Setup and Operation 3 17...
Page 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Page 56: ...Section 3 Setup and Operation 3 27...
Page 58: ...Section 3 Setup and Operation 3 29...
Page 67: ...Section 3 3 38 Setup and Operation...
Page 70: ...Section 3 Setup and Operation 3 41...
Page 107: ...Section 6 6 6 Troubleshooting...