Appendix D
Appendix D
D-7
(EOL "{hex code(s)"})
End of Line character. Enter hex value in double quotes.
Example:
(Outputs(RS232(EOL "0D0A)))lf
would terminate data strings with a carriage return and a line feed.
(Ndx {bool})
Determines whether an index value is transmitted or not.
Example:
(Outputs(RS232(Ndx TRUE)))lf
would cause the data stream to contain an index value. The index value is incremented
approximately every 6.5 milliseconds (e.g. 152 Hz) and ranges from approximately -2.0E8 to
+2.0E8.
(DiagRec {bool})
Controls whether independent (Diagnostic) text records are ever sent. If TRUE, operates
normally (1/second or on a change). If FALSE, never sends a Diagnostic record.
Example:
Transmitted:
(Outputs(RS232(DiagRec TRUE)))lf
Received:
(Diagnostics (SYNC TRUE)(PLL TRUE)(DetOK TRUE)(Chopper TRUE)(Path
61))
(DiagVal {bool})
Controls whether or not a 1 byte diagnostic value (0-255) is output (same as SDM diagnostic
value, see page 3-28) in the (Data stream. "Diag" is used for the label when (Labels TRUE is
set.
Summary of Contents for LI-7500
Page 1: ...Analyzer Environmental Division Instruction Manual 7500...
Page 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Page 41: ...Section 3 3 12 Setup and Operation...
Page 43: ...Section 3 3 14 Setup and Operation...
Page 46: ...Section 3 Setup and Operation 3 17...
Page 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Page 56: ...Section 3 Setup and Operation 3 27...
Page 58: ...Section 3 Setup and Operation 3 29...
Page 67: ...Section 3 3 38 Setup and Operation...
Page 70: ...Section 3 Setup and Operation 3 41...
Page 107: ...Section 6 6 6 Troubleshooting...