Section 3
3-16
Setup and Operation
Cal Coefficients Page
The LI-7500 uses a fifth order polynomial for the CO
2
calibration, and a third order
polynomial for H
2
O calibration. The Calibration Coefficients page displays these factory-
determined calibration coefficients, as well as a factor for correcting CO
2
measurements for
band broadening due to the presence of water vapor, and a zero drift correction factor (Z).
The coefficient shown as XS (Cross Sensitivity) compensates for slight cross sensitivity
between CO
2
and H
2
O signals absorbed by the detector (see Cross Sensitivity on page 2-4,
and Zero Drift on page 2-5).