FCT 76
System Software
5-29
76 Vapor Flag
Water corrections may be selected in the LI-6262 using FCT 76. The Vapor
Flag default setting is for both pressure broadening and dilution corrections
to be applied if a System Reset (FCT 08) is performed. See Sections 3.4 and
3.5 for complete explanations of the vapor corrections.
Corrections Off
↑
,
↓
,
↵
No vapor corrections performed.
State = "0"
Band Broadening
↑
,
↓
,
↵
Correct for band broadening.
State = "1"
BandBrd, Dil
→
REF
↑
,
↓
,
↵
Band broadening & dilution
corrections applied.
State = "2"
*76X
(where x is 0, 1, or 2 for no correction, correct for band broadening,
or correct for band broadening and dilution, respectively).