FCT 16
5-16
System Software
16 Print Calibration Data
Press
FCT 16
to send the current calibration stack for H
2
O, CO
2
, and
Miscellaneous channel codes to the RS-232C port. The output includes the
calibration constants, auxiliary channel equations, and zero and span
parameters for H
2
O and CO
2
. The output should look similar to this:
CO2
T: 3.5970E+01
K: 1.7913E+04
A: 1.4330E-01
B: 9.5609E-06
C: 7.8293E-09
D:-1.1040E-12
E: 7.5366E-17
Ref: 0.0000E+00
Zero: 0.0000E+00
Span: 1.0000E+00
VC:Corrections Off
H2O
T: 4.1340E+01
K: 1.5409E+04
A: 6.3281E-03
B: 3.1059E-06
C: 1.1238E-12
Ref: 0.0000E+00
Zero: 0.0000E+00
Span: 1.0000E+00
MISC
AUX DEST: 43
AUX A: 5.9071E+01
AUX B: 1.5360E-02
AUX C: 0.0000E+00
AVG TIME: 1
PRESSURE: 9.6805E+01
VP CORR A: 1.5000E+00
*16