Console Commands
E-3
Remote Commands
Function
Type the following on a computer or terminal
(comments).
01 Set CO
2
Calibration
02 Set H
2
O Calibration
05 Set D/A Conversion Code #1
06 Set D/A Conversion Code #2
07 Backlight
09 Integrate
11 Print Header
12 Print Data
13 Set Print List
14 Auto Print
15 Auto Header
16 Print Calibration Data
18 Reset Timer
19 Full Status
46 Integration A
47 Integration Peak Value
49 Software Revision I.D.
71 Auxiliary A
72 Auxiliary B
73 Auxiliary Destination
74 Set Average Time
75 Auxiliary C
76 Vapor Flag
77 Pressure
78 H
2
O Cal. a
91 Display 1
|
99 Display 9
*01T,K,A,B,C,D,E,REF,VAPOR FLAG
*02T,K,A,B,C,REF
*05x,low,high
(x is channel code 22-27, 32-38,
42)
*06x,low,high
(same as 05)
*07Y
(Yes) or
*07N
(No) to turn backlight
ON/OFF)
*09x,start,[start value],stop, [stop value]
(x is
channel code 22-29, 32-39)
*11
(prints header)
*12
(prints data)
*13x,x,x,x,x,x,x,x,x,x
(x is any channel code 21-
29, 31-39, 41-47)
*14x
(prints a line of data every x seconds
where x is any number 0.2, 0.5, or 1 to 3600)
*15x
(prints a column header every x lines of
data, from 1 to 999)
*16
(prints calibration stack for H
2
O & CO
2
)
*18
(resets the timer to zero)
*19x
(x is optional parameter to output)
*46
*47
*49
*71<A>
(A is the auxiliary A value)
*72<B>
(B is the auxiliary B value)
*73x
(x is auxiliary destination channel <null>,
29, 39, or 43)
*74x
(x is any value from 0 to 30)
*75<C>
(C is auxiliary C value)
*76x
(x is 0, 1, or 2 to turn ON/OFF
vapor/dilution corrections)
*77x
(x is barometric pressure in kPa)
*78x
(x is vapor correction a)
*91x,x
(x is channel code 21-27, 29, 31-39, 41-
44, for top and bottom line)
(
*91 - *99
are all the same)