![LI-COR li-6262 Instruction Manual Download Page 67](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652067.webp)
FCT 40-45
5-22
System Software
40 Miscellaneous
40 ** MISC **
↑
,
↓
,
↵
Enters menu at function 40.
41-44 Temperature and Auxiliary
Press
FCT
, followed by the appropriate code, to view the temperature and
auxiliary channel codes. These channels can only be viewed by pressing
FCT
; they are not executable from the main menu.
Code
Description
41
Temperature mV
42
Temperature
°
C
43
Pressure (kPa)
44
Auxiliary mV
NOTE
: Pressure (kPa), FCT 43 is used for computations. This comes from
FCT 77, or an analog signal at FCT 44 (see Section 6.4).
45 Timer
Timer 3.7
Displays accumulated time in
seconds from instrument start-
up.
The Timer acts as a counter which accumulates the number of seconds since
instrument start-up (to 999,999), or since the Reset Timer function (FCT 18)
was last executed.
The Timer can also be viewed by entering "45" into any of displays 1 to 9
(FCT 91-99). The Timer function can be output to a computer or printer by
entering "45" into the Set Print List function (FCT 13).