FCT 02-03
System Software
5-5
02 H
2
O Calibration
Function 02 performs functions 61-65, and 68.
H
2
O calibration coefficients (T,K,A,B,C) are factory-entered, and are on the
calibration sheet. H
2
O reference mole fraction is in units of millimoles/mole
of water vapor in the reference cell.
Verify these values with the H
2
O calibration sheet and change, if needed.
Prompt
Description
Related Function
H2 T=
Calibration Temperature
61
H2 K=
Calibration Constant K
62
H2 A=
Calibration Coefficient A
63
H2 B=
Calibration Coefficient B
64
H2 C=
Calibration Coefficient C
65
H2 R=
H
2
O Reference mmol/mol
68
*02T,K,A,B,C,REF
03 Set Miscellaneous
Function 03 performs functions 71-75 and 77.
Prompt
Description
Function
P(kPa)=
Barometric Pressure
77
Avg(s)=
Signal Averaging Time
74
Aux A=
Auxiliary Input Coefficient A
71
Aux B=
Auxiliary Input Coefficient B
72
Aux C=
Auxiliary Input Coefficient C
75
Aux Dest=
Auxiliary Destination
73
There is no remote command for FCT 03, but all of the 7x functions can be
accessed remotely.