GA130001441_002_C0
– 09/2016
43
8.3.2
RPV (Read Pressure Value)
Reading a pressure value for a measurement channel.
S:
RPV[a]<CR>
E:
b[,][TAB]x.xxxxE±xx
Parameter
Description
a
Channel number
1 = Channel 1 (TM1)
2 = Channel 2 (TM2)
3 = Channel 3 (PM)
b
Status
0 = Measured value OK
1 = Measured value < measurement range
2 = Measured value > measurement range
3 = Measured values << measurement range (Err
Lo)
4 = Measured values >
>
measurement range (Err
Hi)
5 = Sensor off (S oFF)
6 = HV on (HU on)
7 = Sensor error (Err S)
9 = No sensor(no Sen)
10 = No switch-on or switch-off threshold (notriG)
12 = Pirani error (Err Pi)
x.xxxxE
±
xx
Pressure value of the selected channel in the
current unit of measurement
8.3.3
RVN (Read Version Number)
Reading the version number of the instrument software.
S:
RVN<CR>
E:
x.xx<CR>
Parameter
Description
x.xx
Versions number