46
GA130001441_002_C0
– 09/2016
Parameter
(continued)
Description
d
Brightness
0 = hoch
1 = gering
e
Profibus
1
– 126
f
Baud rate
0 = 9600
1 = 19200
2 = 38400
g
Serial interface
0 = RS232
1 = RS485
8.3.11
SGP (Set General Parameter)
Setting instrument settings.
S:
SGP[a, b, c, d, e, f, g]<CR>
E:
OK <CR>
Parameter
Description
a
Unit
0 = mbar
1 = Pa
2 = Torr
X = Parameter remains unchanged
b
Analog output
0 = CM 31 mode
1 = CM 51 mode
X = Parameter remains unchanged
c
Number of displayed digits
0 = 2
1 = 3
X = Parameter remains unchanged
d
Brightness
0 = hoch
1 = gering
X = Parameter remains unchanged
e
Profibus
1..126
f
Baud rate
0 = 9600
1 = 19200
2 = 38400
X = Parameter remains unchanged
g
Serial interface
0 = RS232
1 = RS485
X = Parameter remains unchanged
CAUTION:
Changes to the parameters baud rate and Profibus address become only
effective after having restarted the instrument. Before restarting, save the
current configuration using the SAC write command.