Configuration control
Leuze electronic
DCR 50
55
8.2.13 Setup motion detection parameters
Code Description
Command Format Op-
tions
De-
fault
Notes/Examples
Get All motion detect set-
tings
MD PM
G
Returns all motion detection parameter
values in an XML element.
Example
: MDPMG
Minimum Illumination
MD PM
S/P/R/
G
NI
0
0
Minimum Value
Minimum illumination is the lowest value
the AGC should use to set the illumina-
tion.
Valid Range:
0 to Maximum illumination
Example
: MDPMSNI1
Maximum illumination
MD PM
S/P/R/
G
XI
6
100
Maximum Value
This is the highest value the AGC
should use to set the illumination
Valid Range:
Minimum illumination to
100
Example
: MDPMSXI0
Initial illumination value
MD PM
S/P/R/
G
II
1
The starting value the AGC will use to
start adjusting illumination
Valid Range:
Minimum illumination to
Maximum illumination
Example
: MDPMSII1
Minimum exposure time (µs) MD PM
S/P/R/
G
NE
1
1
Minimum Value
Valid Range:
1 to Maximum exposure
time microseconds
Example:
MDPMSNE100
This is the minimum time the camera
lets light into the element to take the pic-
ture in microseconds.
Maximum exposure time
(µs)
MD PM
S/P/R/
G
XE
46
200
00
Maximum Value
Valid Range:
Minimum exposure time
to 20000 microseconds
Example:
MDPMSXE10040
Initial exposure time (µs)
MD PM
S/P/R/
G
IE
40
Valid Range:
Minimum exposure time
to Maximum exposure time microsec-
onds
Example
: MDPMSIE100
Minimum gain
MD PM
S/P/R/
G
NG 1
0
Minimum Value
Valid Range:
0 to Maximum Gain
Example
: MDPMSNG15
Maximum gain
MD PM
S/P/R/
G
XG
47
64
Maximum Value
Gain is the amount of signal amplifica-
tion the AGC can apply to make the pic-
ture easier to read
Valid Range:
Minimum Gain to 64
Example
: MDPMSXG35