Diagnostics and troubleshooting
Leuze electronic
AMS 335
i
89
TNT
35/7-2
4
V
Notice!
Please use
chapter 10 as a master copy
should servicing be required.
Cross the items in the "Measures" column which you have already examined, fill out the fol-
lowing address field and fax the pages together with your service contract to the fax number
listed below.
Customer data (please complete)
Leuze Service fax number:
+49 7021 573 - 199
ATT
(insufficient received signal
level)
Reflector soiled
Clean reflector or glass lens.
Glass lens of the AMS soiled
Performance reduction due to snow, rain, fog, con-
densing vapor, or heavily polluted air (oil mist, dust)
Optimize usage conditions.
Laser spot only partially on the reflector
Check alignment.
Protective foil on the reflector
Remove protective foil from reflector.
TMP
(operating temperature out-
side of specification)
Ambient temperatures outside of the specified
range
In case of low temperatures, remedy may be an
AMS with heating.
If temperatures are too high, provide cooling or
change mounting location.
LSR
Laser diode warning
Laser diode prefailure message
Send in device at next possible opportunity to have
laser diode replaced. Have replacement device
ready.
ERR
Hardware error.
Indicates an uncorrectable error in the hardware
Send in device for repair.
Device type:
Company:
Contact partner / department:
Phone (direct):
Fax:
Street / No:
ZIP code/City:
Country:
Display
Possible error cause
Measure