Technical data
52
Technical data
Power supply:
230 VAC, 115 VAC or 24 VAC
±
10%, 50 -
60 Hz
Instrument fuse 230 V: T0,1 A
Instrument fuse 115 V: T0,2 A
Instrument fuse 24 V: T1,0 A
Power consumption:
max. 30 VA, without external load
Protection class:
I
Degree of protection:
IP 65
Conformity:
EN 61000-6-2, EN 61000-6-4,
EN 61010-1
Ambient temperature:
10
– 45 °C
Measuring range:
See section "Performance specifications"
Current interface:
0/4 - 20 mA, max. load 500 Ohms
Log printer:
See section "Accessories"
Dimensions:
W x H x D = 380 x 480 x 280 mm
Weight:
Approx. 9.5 kg
Other:
The instrument is non-volatile
Water connection
Operating pressure:
1 to 8 bar / 1x10
5
to 8x10
5
Pa
or
0.3* to 1 bar / 0.3x10
5
to 1x10
5
Pa
(Remove flow controller valve body 11225! )
Water inlet:
Opaque pressure hose with external
diameter 6/4x1 mm
Water outlet:
Pressure hose with internal diameter 12
mm
Water temperature:
10 to 40 °C
* When using Testomat 2000
®
at a pre-pressure of 0.3 bar, it must
be ensured that the flow rate through the measuring chamber is at
least 400 ml/min.
We reserve the right to make technical changes without notice in
the interest of constantly improving our products!
Our manuals are updated regularly. If you have an older version
(see version at the back of the manual) you will find the current
manual on our website www.heyl.de on the download page.