Installation
10
Installation
Risks resulting from incorrect installation!
Install Testomat 2000
®
at a location where it is protected against
dripping or splash water, dust and aggressive substances
– e.g. in
a switch cabinet or on a suitable wall.
Information for trouble-free operation
Install Testomat 2000
®
vertically and without mechanical stress.
Install Testomat 2000
®
at a vibration-free site.
Operating Testomat 2000
®
in the pressure
range of 0.3 to 1 bar
Prior to installation, please check whether lower operating pressure is
required. The instrument is factory set for the operating range of 1 to
8 bar. Remove the flow controller valve body
to operate the
instrument in the operating range of 0.3 to 1 bar (e.g. when using an
aerator type R). This involves removing the retaining pin
from the
controller/filter receiver
. Subsequently use the metal bracket to
remove the controller plug
from the borehole. Then remove the
flow controller valve body
and reinsert the controller plug and the
retaining pin.
Installing Testomat 2000
®
Select an installation site where the water inlet hose can be kept as
short as possible (max. 5 m).
Please leave sufficient space on the left-hand side of the
instrument to open the door.
Drill the mounting holes as shown in the drawing on the left.
Use three screws to attach the instrument at a suitable position in
the switch cabinet or on a wall.
Remove for
pressure range
0.3 to 1 bar
WARNING
!
NOTE