Error messages/Troubleshooting
46
Error messages/Troubleshooting
Displayed message (flashes at
selected display)
Instrument result functions
Possible causes
Remedies
Ff. POWER FAILURE 24 V
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Internal power failure of the
24 V supply
Replace fuse F4 or F8
(The control lamp "Power"
of the dosing pump should
illuminate)
Ff. DOSING PUMP
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Dosing pump is defective
- No dosing message from
the dosing pump
Replace dosing pump
Check cable to the dosing
pump for correct
connection
Mf. TURBID
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message
- Continue measurements
- The water is turbid / soiled
MEASURING RANGE EXCEEDED
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message
- Continue measurements
- The measuring range has
been exceeded
Select another type of
indicator (basic program)
LOW WATER LEVEL
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- No water input although
LED
"IN"
lamp illuminates
- Inlet pressure too low
- Overflow detection does not
react
Check water inlet
Connector at the inlet
valve oxidised
Clean filter strainer
Replace valve block
Extract flow controller
valve body
Replace fuse F6
Ff. OUTLET TO DRAIN
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Water remains in the
measuring chamber
although LED "OUT"
illuminates
Check water outlet
Connector at the outlet
valve oxidised
Replace valve block
LOW INDICATOR LEVEL
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message LED
and output "maintenance" on
- Continue measurements
- Minimum indicator quantity
not reached
Without BOB: 50 ml (10%),
With BOB: According to
calculation
Check indicator level and,
if necessary, refill
(enter the filling quantity!)
Mf. SOILING
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses or no message LED
and output "maintenance" on
- Continue measurements
- Sight-glass windows are
soiled
Clean sight-glass windows
Ff. OPTICS
CANCEL WITH HORN KEY
- After programming:
Continuous alarm or message
impulses
- Standby
- Plug-in circuit board
defective
- Error at the optical
component (transmitter or
receiver defective)
Replace plug-in circuit
board
Replace the measuring
chamber holder
Abbreviations: Ff.: = function fault, Mf. = Measuring fault