88
TS11/TS15, Check & Adjust
4.
Meas
to measure and to continue to the next screen.
Motorised instruments change auto-
matically to the other face.
Non-motorised instruments guide to
the other face.
)
The fine pointing must be
performed manually in both faces.
5.
Face II measurement
Meas
to measure the same target in the other face and to calculate the
tilting axis error.
)
If the error is bigger than the predefined limit, the procedure must be
repeated. The tilting axis measurements of the current run are then
rejected and not averaged with the results from previous runs.
Step
Description
180°
180°
TS_071