152
TS11/TS15, Technical Data
Atmospheric condi-
tions
)
Measurements can be made to reflector tapes over the entire range without external
ancillary optics.
Accuracy
Accuracy refers to measurements to standard prisms.
Beam interruptions, severe heat shimmer and moving objects within the beam path
can result in deviations of the specified accuracy.
The display resolution is 0.1 mm.
Range A:
Strong haze, visibility 5 km; or strong sunlight, severe heat shimmer
Range B:
Light haze, visibility about 20 km; or moderate sunlight, slight heat
shimmer
Range C:
Overcast, no haze, visibility about 40 km; no heat shimmer
EDM measuring
mode
std. dev.
ISO 17123-4,
standard prism
std. dev.
ISO 17123-4,
tape
Measurement
time, typical [s]
Standard
1 mm + 1.5 ppm
5 mm + 2 ppm
2.4
Single (fast)
3 mm + 1.5 ppm
5 mm + 2 ppm
0.8
Continuous
3 mm + 1.5 ppm
5 mm + 2 ppm
< 0.15
Average
1 mm + 1.5 ppm
5 mm + 2 ppm
-