background image

12

Version 1.2, Revision C 

 

Operation

5

5. 

Operation

5.1 

Prerequisite for sectioning the specimens

•  To use it with the HistoCore rotary microtomes, a vertical travel range of at least 70mm, and an 

adjustable force balance, are necessary.

•  To use the non-directional specimen fixture or the directional specimen fixture (fine or only XY), bring 

the specimen fixture into the zero position.

•  The embedded tissue samples to be sectioned must be precooled to -5 °C (+23 °F) using ice or a 

cooling plate prior to sectioning.

•  After the RM CoolClamp's precooling phase of 30 min., the universal cassette can be inserted and the 

sections carried out.

•  Permitted maximum sectioning speed: 155 sections/min..
•  Trim at maximum 40μm and section at a thickness of 1-15 μm.

Warning 

•  Operating with larger thicknesses will lead to potential damage of the specimen block.

5.2 

Operating concept

Warning 

•  Installation on a HistoCore rotary microtome, the following applies in general: Handling specimen 

head or knife / blade holder, always lock handwheel in place. Release locking mechanism only for 

sectioning and use safety guard; see Safety information 

(

 P. 7 – 2. Safety)

.

For a microtome that is already configured:

1.  Remove knife / blade holder with base and specimen clamp. 
2.  Bring the specimen head into the rear end position (home position).
3.  Lock the handwheel in place and, where appropriate, check whether the hand wheel lock has 

been activated.

4.  We recommend working with the non-directional specimen holder fixture. If you will be 

working with the directional specimen fixture (XY or fine), it must be brought into the zero 
position.

5.  Push the RM CoolClamp with adapter into the dovetail guide of the quick clamping system for 

specimen clamps and tighten it using a size 4 Allen key.

6.  To carry out the force balancing, please observe the Instructions for Use of the respective 

microtome under the Operation chapter, Fine Adjustment of the Force Balance. The specimen 
head must not fall into the knife after being stopped/released.

7.  Reinstall the knife / blade holder with base and select the desired clearance angle.
8.  Attach the cable holder to the microtome; cable holders are included in the standard delivery 

of the RM CoolClamp.

5.3 

Installing the RM CoolClamp to Rotary Microtomes with Quick Clamping System

Tip 

The RM CoolClamp is attached the same way as changing the universal cassette clamp; for 

additional information, refer to the Instructions for Use of the respective rotary microtomes.

Summary of Contents for RM CoolClamp

Page 1: ...lly Cooled Universal Cassette Clamp Instructions for Use English Order No 14 0502 82101 Revision C Always keep this manual with the instrument Read carefully before working with the instrument Version...

Page 2: ......

Page 3: ...l details of the present Instructions for Use are not to be considered warranted characteristics of our products These are determined only by the contract provisions agreed between ourselves and our c...

Page 4: ...ctions 9 4 2 1 Ventilation openings 9 4 2 2 Position of the cable duct 10 4 2 3 Potential pinching areas 10 4 2 4 Power pack with EU UK UL AU and CCC adapters 11 5 Operation 12 5 1 Prerequisite for se...

Page 5: ...ic equipment in accordance with Section 7 of the German Electrical and Electronic Equipment Act ElektroG ElektroG is the law on the bringing into circulation return and environmentally compatible disp...

Page 6: ...indicated on the title page A nameplate indicating the instrument serial number is attached at the back side of the instrument 1 3 Intended use The RM CoolClamp is an electric cooling specimen holder...

Page 7: ...t must not be operated in rooms with danger of explosion The RM CoolClamp must be operated only in the installation orientation and position prescribed for it It must be ensured that no liquids ingres...

Page 8: ...7 5 V DC Additional data for the RM CoolClamp Operating temperature 10 C to 35 C Operating relative humidity 10 to 80 non condensing Transport temperature 29 C to 50 C Storage temperature 5 C to 50 C...

Page 9: ...omes 4 2 General operating instructions 4 2 1 Ventilation openings Warning To prevent injury and damage provision must be made for an unobstructed air supply refer also to the cleaning instructions in...

Page 10: ...p must be routed to prevent possible severing of the cable when operating the microtome For this purpose attach the cable holder to the microtome and then insert the cable into the cable duct Fig 3 1...

Page 11: ...L AU and CCC adapters Fig 5 1 Select the appropriate adapter for the power supply If the provided plugs of the adapter do not fit to your local power supply please use a converting plug to work with o...

Page 12: ...r a microtome that is already configured 1 Remove knife blade holder with base and specimen clamp 2 Bring the specimen head into the rear end position home position 3 Lock the handwheel in place and w...

Page 13: ...d position by turning the handwheel Fig 6 2 and engage the handwheel lock 2 To release the clamping system turn the screw Fig 6 3 counterclockwise using a size 4 Allen key 3 Push the guide Fig 6 5 of...

Page 14: ...correct If the specimen head moves i e it is raised or lowered fine adjustment is necessary Warning Failure to adjust the force balance may result in injury while working see the Safety information P...

Page 15: ...ventilation openings are free before and during use refer to P 9 4 2 1 Ventilation openings No cleaning in the incubator is permitted The RM CoolClamp must not be disassembled for cleaning 6 2 Mainte...

Page 16: ...w LeicaBiosystems com within the product menu This template has to be used for gathering all required data When returning a product a copy of the filled and signed confirmation has to be enclosed or p...

Page 17: ......

Page 18: ...w LeicaBiosystems com Version 1 2 Revision C 11 2019 Leica Biosystems Nussloch GmbH Heidelberger Str 17 19 D 69226 Nussloch Germany Tel 49 0 6224 143 0 Fax 49 0 6224 143 268 Web www LeicaBiosystems co...

Reviews: