Operation
5
Fig. 7
1
2
If another accessory of a different weight is mounted on the specimen head
check whether it is necessary to readjust the force balance.
Checking the correct setting:
• Attach a new accessory and clamp a specimen.
• Set the specimen head to half the height of the vertical travel range by turning the handwheel
If the specimen head remains in this exact position, the setting is correct.
If the specimen head moves, i.e. it is raised or lowered, fine adjustment is necessary.
Warning
• Failure to adjust the force balance may result in injury while working; see the Safety information
.
The force balance is adjusted using the screw
, which can be accessed by removing the
section waste tray on the bottom of the base plate of the microtome. Use the Allen key provided, size 5
(with handle!) for the adjustment.
• If the specimen head moves downwards, turn the screw
clockwise.
• If the specimen head moves upwards, turn the screw
counterclockwise.
• Continue this procedure until the specimen head no longer moves once released.