RM CoolClamp

 

Setting up the instrument

4

4. 

Setting up the instrument

4.1 

Standard delivery - packing list

Quantity

Part description

1

RM CoolClamp

14 0502 46573

1

Power pack with EU, UK, UL, AU and CCC adapters

14 6000 05329

1

Cable holders for attaching to the microtome

14 6000 05334

1

Instructions for Use (printed English with language CD)

14 0502 82001

Fig. 1 

 

Tip 

The RM CoolClamp can be used only together with the quick clamping system on the HistoCore rotary 

microtomes.

4.2 

General operating instructions

4.2.1  Ventilation openings

Warning 

•  To prevent injury and damage, provision must be made for an unobstructed air supply; refer also 

to the cleaning instructions in the Instructions for Use 

(

 P. 15 – 6. Cleaning and Maintenance)

The fan must not be blocked and the air outlet must not be obstructed. Please check before each 

use that the fan rotates freely. To prevent damage to the RM CoolClamp, the maximum permissible 

ambient temperature must not be exceeded; for additional information, refer to the Technical data 

(

 P. 8 – 3. Technical Data)

.

Tip 

When operating the RM CoolClamp , condensation on the clamp occurs. This is a natural physical 

process.

Summary of Contents for RM CoolClamp

Page 1: ...lly Cooled Universal Cassette Clamp Instructions for Use English Order No 14 0502 82101 Revision C Always keep this manual with the instrument Read carefully before working with the instrument Version...

Page 2: ......

Page 3: ...l details of the present Instructions for Use are not to be considered warranted characteristics of our products These are determined only by the contract provisions agreed between ourselves and our c...

Page 4: ...ctions 9 4 2 1 Ventilation openings 9 4 2 2 Position of the cable duct 10 4 2 3 Potential pinching areas 10 4 2 4 Power pack with EU UK UL AU and CCC adapters 11 5 Operation 12 5 1 Prerequisite for se...

Page 5: ...ic equipment in accordance with Section 7 of the German Electrical and Electronic Equipment Act ElektroG ElektroG is the law on the bringing into circulation return and environmentally compatible disp...

Page 6: ...indicated on the title page A nameplate indicating the instrument serial number is attached at the back side of the instrument 1 3 Intended use The RM CoolClamp is an electric cooling specimen holder...

Page 7: ...t must not be operated in rooms with danger of explosion The RM CoolClamp must be operated only in the installation orientation and position prescribed for it It must be ensured that no liquids ingres...

Page 8: ...7 5 V DC Additional data for the RM CoolClamp Operating temperature 10 C to 35 C Operating relative humidity 10 to 80 non condensing Transport temperature 29 C to 50 C Storage temperature 5 C to 50 C...

Page 9: ...omes 4 2 General operating instructions 4 2 1 Ventilation openings Warning To prevent injury and damage provision must be made for an unobstructed air supply refer also to the cleaning instructions in...

Page 10: ...p must be routed to prevent possible severing of the cable when operating the microtome For this purpose attach the cable holder to the microtome and then insert the cable into the cable duct Fig 3 1...

Page 11: ...L AU and CCC adapters Fig 5 1 Select the appropriate adapter for the power supply If the provided plugs of the adapter do not fit to your local power supply please use a converting plug to work with o...

Page 12: ...r a microtome that is already configured 1 Remove knife blade holder with base and specimen clamp 2 Bring the specimen head into the rear end position home position 3 Lock the handwheel in place and w...

Page 13: ...d position by turning the handwheel Fig 6 2 and engage the handwheel lock 2 To release the clamping system turn the screw Fig 6 3 counterclockwise using a size 4 Allen key 3 Push the guide Fig 6 5 of...

Page 14: ...correct If the specimen head moves i e it is raised or lowered fine adjustment is necessary Warning Failure to adjust the force balance may result in injury while working see the Safety information P...

Page 15: ...ventilation openings are free before and during use refer to P 9 4 2 1 Ventilation openings No cleaning in the incubator is permitted The RM CoolClamp must not be disassembled for cleaning 6 2 Mainte...

Page 16: ...w LeicaBiosystems com within the product menu This template has to be used for gathering all required data When returning a product a copy of the filled and signed confirmation has to be enclosed or p...

Page 17: ......

Page 18: ...w LeicaBiosystems com Version 1 2 Revision C 11 2019 Leica Biosystems Nussloch GmbH Heidelberger Str 17 19 D 69226 Nussloch Germany Tel 49 0 6224 143 0 Fax 49 0 6224 143 268 Web www LeicaBiosystems co...

Reviews: