13
Lamphousing 106 z
*
Hg- and Xe lamps
Attention:
Danger: the following information is
extremely imp ortant and should be adhered
to under all circumstances:
Always unp lug the p ower unit from the mains
before assembly work is carried out.
Wait for the lamp housing to cool down before
op ening (at least 15 min.), danger of ex-
p losion!
Never touch glass p arts of the burner with
your hands. Remove any fingerprints or dust
carefully (perhaps using alcohol).
Adjust lamp s immediately after ignition (see
page 90 ff.)
Attention:
Avoid switching on and off fre quently, as this can
impair the stability of the lamp and shorten its
life.
Hot Hg lamps cannot be reignited until they have
cooled down. We recommend that you let new
burners burn in for several hours without
interruption if possible.
It is a good idea to keep a record of the hours
the lamp is in use and to compare with the
manufacturer’s specifications. Replace dis-
coloured, spent lamps.
We cannot accept any liability for damage
resulting from a lamp explosion.
Attention:
Always wear safety clothing (gloves and face
mask) when assembling Xe burners (danger
of explosion).
Fig. 5
Lamphousing 106 z*
1
Lid, flipped up,
2
Collector,
3
12 V 100 W halogen lamp with
holder
4, 9
Lid screws,
5
Reflector,
6, 8
x/y centering of
reflector,
7
Reflector focusing,
10
Screws for lamp socket,
11
Socket for cut-out plug
Fig. 6
12 V 100 W lamp holder (LH 106 z only)
1
2
3
4
5
6
7
8
9
10
10
11
!