Pol object guide
*
Move the object guide until the fixing screw can
be seen under the drill hole (13.1). Insert the
object guide in the guide holes of the rotary
stage and tighten the fixing screw with the
hexagonal screwdriver.
Attachable object guide
*
The attachable object guide can be fixed on the
left, right or at the front (not illustrated) with the
two clamp screws.
Condenser holder
*
The microscope stage must be equipped with
the condenser holder (12.10) for transmitted light
work. The condenser holder enables various
condensers to be changed quickly and centered
and takes components for polarized light
(Figs. 27.6 and 57.1). An adjustable upper stop
(12.5) guarantees a reproducible vertical setting
of the condenser (Koehler illumination).
Interchangeable stages only: to assemble the
condenser holder, either remove the stage or
move it as far upwards as possible. Loosen the
clamp screw (12.2) slightly with the 3 mm hex-
agonal screwdriver, slide the condenser holder
onto the guide pin and retighten the clamp
screw (12.2) (already assembled for fixed
mechanical stage).
Important! Do not mount at an angle, note the
stop!
Fig. 13
Rotary Pol stage* and Pol 3 object guide*
1
Drill hole for fixing screw,
2
Swing-in/out lever to hold
specimen slides of different formats,
3
Place to keep center-
ing keys,
4
Pairs of clickstop buttons,
5
45
°
clickstop,
6
Stage
rotation clamp