12
Fig. 4
Lamphousing 106*, opened
1
12 V 100 W halogen lamp in holder,
2
Collector,
3
Diffusing
screen
Lamphousing 106 z
Important:
For incident light only (48.1)! Disassembled like
lamphousing 106 (see above).
12 V 100 W halogen lamp
Disconnect from power supply (2.5).
Loosen screws (5.4 and 5.9) with crosstip
screwdriver and flip up lid (5.1).
Pull cut-out plug slightly out of socket (5.11).
Unscrew screws (5.10) on the lamp holder and
pull out the lamp holder (Fig. 6). Remove defect
lamp and insert new 12 V 100 W halogen lamp.
Attention:
Leave the p rotective covering on the lamp until it
is in its holder!
Avoid making fingerprints or wipe off
immediately.
Spare lamps
See page 112 for code nos.
Lamphousing 106
Disconnect from power supply (2.5),
disassemble using hexagonal screwdriver (1.1
and 3.2). Unscrew screw (2.9) and remove cover.
Move the collector to the front (48.19).
Remove the defect lamp and put a new
12 V 100 W halogen lamp into the lamp holder
without tilting (4.1).
Attention:
Leave the protective covering on the lamp until it
is in its holder. Avoid making fingerprints on the
lamp or wipe off immediately.
Close the lamphousing (2.9).
!
1
2
3
!
!