Leica CM1510 – Cryostat
7
Cleaning and disinfection
• Do not use organic solvents or any other
aggressive substances for cleaning and
disinfection!
We strongly recommend the use of Leica
Cryofect disinfectant spray! Only use al-
cohol or common disinfectants based on
alcohol!
• For spray disinfection follow the instruc-
tions for use supplied with the disinfec-
tant!
• Do not use external heaters for drying the
cryochamber. This can cause severe dam-
age to the cooling system!
Removal of the microtome
• Before removing the microtome, turn the
instrument off with the mains switch and
disconnect it from mains!
• Before taking the microtome out of the
cryochamber, place the handle of the
handwheel in the lowest position.
When removing the microtome, the speci-
men head will rapidly fall down and might
injure the operator‘s hands!
• Wear appropriate protective gloves to re-
move the cold microtome from the cryo-
chamber. Risk of frost bite!
• The microtome must be completely dry be-
fore reinstallation. Humidity inside will
condense and freeze in the cold cryostat
and thus may cause malfunctions or dam-
age.
3.
Safety instructions for handling the instrument
Maintenance
• Only qualified and authorized service per-
sonnel may access the internal compo-
nents of the instrument for service and re-
pair.
• Clean the compressor ventilation grid at
least once a year or more often if required!
Fuse replacement (Chapter 11)
• Turn the instrument off with the mains
switch and pull the mains plug, before re-
placing the fuses.
•
Only use fuses of the same specification!
For the required values, please refer to
Chapter 4 'Technical data'.
Lamp replacement (Chapter 11)
• Turn the instrument off with the mains
switch and pull the mains plug, before re-
placing the lamp.
• If the lamp is broken, it must be replaced
by the technical service, as the replace-
ment involves a high risk of injury.
• Do not use lamp type other than specified
by the manufacturer!