Leica CM1510 – Cryostat
11
The instrument must be transported in an up-
right position, or at an angle of max. 30°!
Do not grip the cabinet at the lid. Grip the cabi-
net only at the marked areas (
), which are
reinforced for transporting (Fig. 11).
•
Screw the adjustable feet
(1)
to the highest posi-
tion by means of a fork wrench (size 16).
•
Move the instrument to the installation site on its
rollers
(2 and 3)
.
•
The adjustable feet
(1)
can support the weight of
the instrument when tipping at a
slight
angle
(max. 30°).
•
To lift the instrument, please use the handle straps
provided, which can be attached to the front roll-
ers
(2)
and rear rollers
(3)
.
•
At the installation site unscrew the adjustable feet
(1)
with a fork wrench (size 16) to ensure stability.
•
Align the adjustable feet to level the instrument.
Do not operate in rooms with explosion haz-
ard!
To ensure an adequate cooling capacity, the
instrument has to be set up with at least 10
cm distance from walls and furniture.
Do not place anything next to the compressor
ventilation grid (right side of the cabinet) to
ensure adequate ventilation!
The place of installation must meet the following re-
quirements:
-
No direct sunlight
-
Mains power socket at a distance no greater than
the length of the power cord (length approximately
2.5 meters) - do not use an extension cord
-
No drafts (air condition outlets, etc.) directly over
the instrument
-
Even floor
-
Mainly vibration-free floor
-
Obstruction-free access to the handwheel
-
Room temperature always below 22 °C
-
Air humidity must not exceed 60%
High room temperatures and excessive air
humidity affect the cooling capacity of the cry-
ostat.
6.
Unpacking and installation
6.1
Site requirements
6.2
Transport and relocation
Rear
Fig. 11
2
3
1