T.O. 33K3-4-3760-1
7
VERTICAL
3
Offset
750
mV
Averaging
5
, then
OK
VERTICAL
4
Offset
750
mV
Averaging
5
, then
OK
Measure Measure
Setup
Show
Table
(checked)
P1
Mean
C1
P2
Mean
C2
P3
Mean
C3
P4
Mean
C4
Vertical
2
(press twice to turn channel off)
Vertical
3
(press twice to turn channel off)
Vertical
4
(press twice to turn channel off)
Close
4.2.2 Connect the Oscilloscope Calibrator to the TI 1 connector.
4.2.3 Set the Oscilloscope Calibrator Amplitude for -750 mV DC into 1 M
. Set the OUTPUT to ON.
4.2.4 Press TI
Clear Sweeps
and allow at least 100 sweeps as indicated on the TI display before proceeding.
4.2.5 Verify the TI P1:mean(C1) indication is within -764.25 to -735.75 mV.
4.2.6 Touch TI
C1
, set the
Offset
to
-750 mV
, then touch
Close.
4.2.7 Set the Oscilloscope Calibrator Amplitude for +750 mV DC.
4.2.8 Press TI
Clear Sweeps
and allow 100 sweeps as indicated on the TI display before proceeding.