T.O. 33K3-4-3760-1
13
4.4.20 Set the TI controls as follows:
Trigger
Trigger Setup
Slope
Positive
Source
C3
Close
4.4.21 Set the Oscilloscope Calibrator OUTPUT to ON.
4.4.22 Repeat steps 4.4.4 through 4.4.12 for TI C3 using C3 controls and settings.
4.4.23 Disconnect the Oscilloscope Calibrator from the TI 3 connector and connect to the TI 4 connector.
4.4.24 Press TI Vertical
3
, then Vertical
4
to turn C3 off and C4 on.
4.4.25 Set the TI controls as follows:
Trigger
Trigger Setup
Slope
Positive
Source
C4
Close
4.4.26 Set the Oscilloscope Calibrator OUTPUT to ON.
4.4.27 Repeat steps 4.4.4 through 4.4.12 for TI C4 using C4 controls and settings.
4.4.28 Disconnect the test setup.
4.5 TIME BASE ACCURACY CALIBRATION:
4.5.1 Set the TI controls as follows:
File
Recall
Setup…
Recall Default
Horizontal
s
ns 500
ms
Vertical
2
(press twice to turn channel off)
C1
Volts/div
100
mV
Coupling
50
Ω
DC