T.O. 33K3-4-3760-1
10
4.3.3 Set the Oscilloscope Calibrator for a 600 mV p-p leveled sine wave at 50 kHz into 50
. Set the OUTPUT to
ON.
4.3.4 Adjust TI Trigger
Level
as necessary for a stable display.
4.3.5 Adjust the Oscilloscope Calibrator output controls for a TI P1:pkpk(C1) indication of as close as possible to
600 mV.
4.3.6 Press TI
Clear Sweeps
and allow at least 100 sweeps as indicated on the TI display.
4.3.7 Set the Oscilloscope Calibrator Frequency to 400 MHz.
4.3.8 Adjust the TI Horizontal controls as necessary for several cycles of display. Touch TI
Timebase
,
Horizontal
Setup
, set the Sampling Mode to
RIS
, then touch
Close
.
4.3.9 Press TI
Clear Sweeps
and allow at least 100 sweeps as indicated on the TI display.
4.3.10 Verify the TI P1:pkpk(C1) average indication is
425 mV.
4.3.11 Set the Oscilloscope Calibrator OUTPUT to OFF.
4.3.12 Disconnect the Oscilloscope Calibrator from the TI 1 connector and connect to the TI 2 connector.
4.3.13 Press TI Vertical
1
, then Vertical
2
to turn C1 off and C2 on.
4.3.14 Touch TI
Timebase
,
Horizontal Setup
,
RealTime
, then set the
Time/Division
to
20.0 µs
.
4.3.15 Touch TI
Trigger
,
Trigger Setup
,
Source
,
C2
,
then
Close
.
4.3.16 Repeat steps 4.3.3 through 4.3.11 for TI C2, using C2 controls, settings and readouts.
4.3.17 Disconnect the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector.
4.3.18 Press TI Vertical
2
, then Vertical
3
to turn C2 off and C3 on.
4.3.19 Touch TI
Timebase
,
Horizontal Setup
,
RealTime
, then set the
Time/Division
to
20.0 µs
.
4.3.20 Touch TI
Trigger
,
Trigger Setup
,
Source
,
C3,
then
Close
.
4.3.21 Repeat steps 4.3.3 through 4.3.11 for TI C3, using C3 controls, settings and readouts.
4.3.22 Disconnect the Oscilloscope Calibrator from the TI 3 connector and connect to the TI 4 connector.
4.3.23 Press TI Vertical
3
, then Vertical
4
to turn C3 off and C4 on.
4.3.24 Touch TI
Timebase
,
Horizontal Setup
,
RealTime
, then set the
Time/Division
to
20.0 µs
.
4.3.25 Touch TI
Trigger
,
Trigger Setup
,
Source
,
C4
, then
Close
.
4.3.26 Repeat steps 4.3.3 through 4.3.11 for TI C4, using C4 controls, settings and readouts.
4.3.27 Disconnect the test setup.