T.O. 33K3-4-3760-1
3
4 CALIBRATION PROCESS:
NOTE
Unless otherwise specified, verify the results of each test and take corrective
action whenever the test requirement is not met, before proceeding.
4.1 DC ACCURACY CALIBRATION:
4.1.1 Set the TI controls as follows:
File Recall
Setup…
Recall Default
Horizontal
s
ns 1.00
s
C1
Volts/div
2.00
mV
Bandwidth
20MHz
Averaging
5
, then
OK
C2
Volts/div
2.00
mV
Bandwidth
20MHz
Averaging
5
, then
OK
Vertical
3
Volts/div
2.00
mV
Bandwidth
20MHz
Averaging
5
, then
OK
Vertical
4
Volts/div
2.00
mV
Bandwidth
20MHz
Averaging
5
, then
OK
Measure Measure
Setup…
Show
Table
(checked)
P1
Mean
C1