![LeCroy WAVESURFER 44MXS-B Technical Manual Download Page 5](http://html1.mh-extra.com/html/lecroy/wavesurfer-44mxs-b/wavesurfer-44mxs-b_technical-manual_1866377005.webp)
T.O. 33K3-4-3760-1
4
P2
Mean
C2
P3
Mean
C3
P4
Mean
C4
Vertical
2
(press twice to turn channel off)
Vertical
3
(press twice to turn channel off)
Vertical
4
(press twice to turn channel off)
Close
4.1.2 Connect the Oscilloscope Calibrator to the TI 1 connector.
4.1.3 Set the Oscilloscope Calibrator for the first value listed in the Applied column of Table 2 into 1 M
. Set the
OUTPUT to ON.
Table 2.
Volts/Div
Applied
(VDC) Limits
(VDC)
2.00 m
+6 m
+4.630 to +7.370 m
-6 m
-7.370 to -4.630 m
5.00 m
+15 m
+13.150 to +16.850 m
-15 m
-16.850 to -13.150 m
10.0 m
+30 m
+27.75 to +32.25 m
-30 m
-32.25 to -27.75 m
20.0 m
+60 m
+56.50 to +63.50 m
-60 m
-63.50 to -56.50 m
50.0 m
+150 m
+142.75 to +157.25 m
-150 m
-157.25 to -142.75 m
100 m
+300 m
+286.5 to +313.5 m
-300 m
-313.5 to -286.5 m