T.O. 33K3-4-3760-1
12
Vertical
4
(press twice to turn channel off)
Close
4.4.2 Connect the Oscilloscope Calibrator to the TI 1 connector.
4.4.3 Set the Oscilloscope Calibrator for a 10 kHz, 700 mV p-p leveled sinewave into 50
. Set the OUTPUT to
ON.
4.4.4 Adjust the Oscilloscope Calibrator for 7.2 div of vertical deflection on the TI display.
4.4.5 Press TI
Clear Sweeps
and allow at least 10 sweeps as indicated on the TI display.
4.4.6 Verify the TI C1
↓
and
↑
(cursors) indications are within -34.0 to +34.0 mV.
4.4.7 Touch TI
Trigger
,
Trigger Setup
, set the slope
to
Negative
, then touch
Close
.
4.4.8 Repeat steps 4.4.5 and 4.4.6.
4.4.9 Touch TI
Trigger
,
Trigger Setup
, set the slope
to
Positive
,
Level
to +250 mV, then touch
Close
.
4.4.10 Repeat steps 4.4.5 through 4.4.8 verifying the TI C1
↓
and
↑
(cursors) indications are 216 to
+284 mV when performing step 4.4.6.
4.4.11 Repeat steps 4.4.9 and 4.4.10 for a TI trigger level setting of -250 mV, verifying the C1
↓
and
↑
(cursors)
indications are within -284 to -216 mV.
4.4.12 Set the Oscilloscope Calibrator OUTPUT to OFF.
4.4.13 Disconnect the Oscilloscope Calibrator from the TI 1 connector and connect to the TI 2 connector.
4.4.14 Press TI Vertical
1
, then Vertical
2
to turn C1 off and C2 on.
4.4.15 Set the TI controls as follows:
Trigger
Trigger Setup
Slope
Positive
Source
C2
Close
4.4.16 Set the Oscilloscope Calibrator OUTPUT to ON.
4.4.17 Repeat steps 4.4.4 through 4.4.12 for TI C2 using the C2 controls and settings.
4.4.18 Disconnect the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector.
4.4.19 Press TI Vertical
2
, then Vertical
3
to turn C2 off and C3 on.