![LeCroy PXA125 User Manual Download Page 31](http://html1.mh-extra.com/html/lecroy/pxa125/pxa125_user-manual_1866284031.webp)
User Manual PXA125
Installation 2-17
Operating
Environment
The PXA125 is intended for operation within a PXI chassis as a
plug-in module. Ensure the PXI chassis being used to host the PXA
125 fully conforms to the latest PXI specifications.
The PXA125 is intended for indoor use and should be operated in a
clean, dry environment with an ambient temperature within the range
of 0 °C to 40 °C.
WARNING
The PXA125 must not be operated in explosive, dusty, or wet
atmospheres. Avoid installation of the module close to strong
magnetic fields.
The design of the PXA125 has been verified to conform to EN
61010-1 safety standard per the following limits:
Installation (Overvoltage) Category I (Measuring terminals)
Pollution Degree 2
Installation (Overvoltage) Category I refers to signal level, which is
applicable for equipment measuring terminals that are connected to
source circuits in which measures are taken to limit transient
voltages to an appropriately low level.
Pollution Degree 2 refers to an operating environment where
normally only dry non-conductive pollution occurs. Occasionally a
temporary conductivity caused by condensation must be expected
The PXA125 operates from within a PXI chassis. DC Voltages are
supplied to the instrument from the PXI backplane. The instrument
requires a variety of DC voltages as outlined in the Specifications
section (Appendix A). Ensure the PXI chassis is capable of
delivering required voltages and has sufficient current to drive the
generator.
Power
Requirements
CAUTION
Disconnect power to the PXI Chassis before installing or removing
the PXA125.
Grounding
Requirements
To conform to the applicable safety and EMC requirements, ensure
that the PXA125 instrument panel and the PXI chassis are “earth”
grounded.
Summary of Contents for PXA125
Page 14: ...User Manual PXA125 xiv This page intentionally left blank ...
Page 58: ...User Manual PXA125 2 44 Installation Figure 2 30 Communicating with your Instrument ...
Page 118: ...User Manual PXA125 3 104 ArbConnection This page intentionally left blank ...
Page 120: ...User Manual PXA125 Programming Reference 4 2 ...
Page 161: ...User Manual PXA125 Programming Reference 4 43 ...
Page 167: ...User Manual PXA125 Programming Reference 4 49 MAXimum 10 in units of volts ...
Page 179: ...User Manual PXA125 Programming Reference 4 61 ...
Page 186: ...User Manual PXA125 Programming Reference 4 68 ...
Page 193: ...User Manual PXA125 Programming Reference 4 75 ...
Page 208: ...User Manual PXA125 Programming Reference 4 90 This page intentionally left blank ...
Page 219: ...User Manual PXA125 Specifications A 101 This page intentionally left blank ...