Minimum voltage swing
The minimum high voltage level is user definable by the hysteresis control up to 1.4 V above the
threshold. The maximum low voltage level is user definable by the hysteresis control up to 1.4 V
below the threshold. The minimum signal swing is 100 mV. The
indeterminate range
of 50 mV
around the threshold voltage level is the level below which the MS-500 will not operate. However,
the MS-500
can
support a signal as low as 100
mV only if the input signal’s quality is adequate.
Basic Oscilloscope Operation
For information on using the oscilloscope to view analog signals, reference your oscilloscope on-
line Help system.
Threshold from - 50 mV to - 1.4 V
Threshold from + 50 mV to + 1.4 V