C
–
6
Appendix C
Parameter
—
Description
Definition
Notes
c s d e v
Cyclic standard deviation:
Standard
deviation of data values from mean value
over integral number of periods. Contrary
to
sdev
, calculation performed over
integral number of cycles, eliminating bias
caused by fractional intervals.
=
−
N
i
i
)
mean
v
(
N
1
2
1
Where: v
i
denotes measured
sample values, and
N =
number
of data points within the periods
found up to maximum of 100
periods.
data
Returns average of all data points.
All data values in
analyzed region
(See Fig. C
–2)
Multi-value parameter especially
valuable for histograms and
trends.
delay
Time from trigger to transition: Measures
time between trigger and first 50 %
crossing after left cursor. Can measure
propagation delay between two signals by
triggering on one and determining delay
of other.
Time between
trigger and first
50 % crossing
after left cursor
(See Fig. C
–2)
dly
delay: Computes time between 50 %
level of two sources.
Time between
midpoint
transition of two
sources
t@lv
t at level: Computes transition between
selected levels or sources.
Time between
transition levels
of two sources,
or from trigger to
transition level of
a single source
Reference levels and edge-
transition
polarity
can
be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
c2d
clock to data
: Computes difference in
time from clock threshold crossing to either
the next (
c2d
+
) or previous (
c2d
−
) data
threshold crossing.
Time from clock
threshold
crossing to next
or previous edge
(See Fig. C
–3)
Threshold levels of clock and
data signals, and edge-transition
polarity
can
be
selected.
Hysteresis argument used to
differentiate peaks from noise in
data, with good hysteresis value
between half expected peak
–to–
peak value of signal and twice
expected peak
–to–peak value of
noise.
Summary of Contents for LSA1000
Page 2: ...Operator s Manual LeCroy LSA1000 Signalyst Revision B October 1998 ...
Page 6: ...iii Contents ...
Page 9: ...v Foreword ...
Page 14: ...1 1 Read This First 1 ...
Page 28: ......
Page 29: ...2 1 2 About the LSA1000 ...
Page 34: ...3 1 3 Installation and Safety ...
Page 42: ...4 8 Connecting to PC or Network Example of responses for IDN and CONET queries ...
Page 43: ...4 1 4 Connecting to PC or Network ...
Page 55: ...5 12 LSA1000 Software Tools Remote LSA display of LSA1000 s External Reference Out Waveform ...
Page 63: ...5 20 LSA1000 Software Tools ...
Page 66: ...5 1 5 LSA1000 Software Tools ...
Page 72: ......
Page 73: ...A 1 A Appendix A Specifications ...
Page 83: ......
Page 84: ...B 1 Appendix B NET_CON Source Code B ...
Page 96: ......
Page 97: ...C 1 C Appendix C Parameter Measurement ...
Page 104: ...D 1 D Appendix D Program Examples ...
Page 107: ...E 1 E Appendix E Suggestion Forms ...
Page 108: ...Index ...
Page 109: ......
Page 114: ...Index ...