5
–
11
Performance Verification with
Remote LSA
Remote LSA
is the simplest way to run the LSA1000’s basic
operations
— without the need for remote control
commands. Thus we recommend that Remote LSA be used
for the initial performance verification of the LSA1000.
1. Ensure that your PC has the Ethernet connection
established with the LSA1000 (see Chapter 4).
2.
Connect the LSA1000’s “REF OUT” to “CH 1”.
3. Start Remote LSA.
4.
In the “Timbase + Trigger” panel click on
to set
the trigger to automatic.
5. Select channel
from the “Channel” panel.
6. The program should display a 10MHz square waveform
of approximately 400mV amplitude, as shown on the
next page.
This process using
Remote LSA
is a simple means of verifiing
that the LSA1000 is working properly when first received.
Remote LSA
is also recommended as a visual diagnostic tool
during LSA1000 program development.
Summary of Contents for LSA1000
Page 2: ...Operator s Manual LeCroy LSA1000 Signalyst Revision B October 1998 ...
Page 6: ...iii Contents ...
Page 9: ...v Foreword ...
Page 14: ...1 1 Read This First 1 ...
Page 28: ......
Page 29: ...2 1 2 About the LSA1000 ...
Page 34: ...3 1 3 Installation and Safety ...
Page 42: ...4 8 Connecting to PC or Network Example of responses for IDN and CONET queries ...
Page 43: ...4 1 4 Connecting to PC or Network ...
Page 55: ...5 12 LSA1000 Software Tools Remote LSA display of LSA1000 s External Reference Out Waveform ...
Page 63: ...5 20 LSA1000 Software Tools ...
Page 66: ...5 1 5 LSA1000 Software Tools ...
Page 72: ......
Page 73: ...A 1 A Appendix A Specifications ...
Page 83: ......
Page 84: ...B 1 Appendix B NET_CON Source Code B ...
Page 96: ......
Page 97: ...C 1 C Appendix C Parameter Measurement ...
Page 104: ...D 1 D Appendix D Program Examples ...
Page 107: ...E 1 E Appendix E Suggestion Forms ...
Page 108: ...Index ...
Page 109: ......
Page 114: ...Index ...