![LeCroy HDO6000 Operator'S Manual Download Page 79](http://html1.mh-extra.com/html/lecroy/hdo6000/hdo6000_operators-manual_1866315079.webp)
Operator's Manual
8. Check
Trace On
to display the new math trace.
Enable/Disable Math Function
Once a math function has been created and saved in one of the F
x
locations, just use the main Math
dialog to quickly enable/disable it.
Touch the Front Panel
Math button
, or from the menu bar, choose
Math > Math Setup
, then check the
On
box next to each function you wish to display.
Clear the On box to disable the function and close the trace.
List of Math Functions
Standard math functions are listed below alphabetically.
NOTE
: There may be additional math functions available depending on the software options installed on
the oscilloscope.
Function
Definition
Absolute
For every point in the waveform the distance away from zero is calculated. For values greater
than zero this is the same as the value. For values less than zero, the magnitude of this value
without regard to its sign is used.
Average
Calculates either a summed or continuous average of a selected number of sweeps. See
Averaging Waveforms
. The maximum number of sweeps is determined by the oscilloscope
model and memory. See the specifications at teledynelecroy.com.
Copy
Copies waveform in its unprocessed state to the first available memory location.
Correlation
Calculates a measure of similarity of two waveforms, or a waveform against itself, as a function
of a time-lag applied to one of them.
Derivative
Calculates the derivative of adjacent samples using the formula:
(next sample value – current sample value) / (horizontal sample interval)
Deskew
Shifts trace in time the amount of the deskew factor.
DIfference
For every point in the waveform, the value of Source2 is subtracted from the value of Source1.
Source1 and Source2 must have the same horizontal units and scale and the same vertical
units.