AP034-OM-E Rev D
ISSUED: January 2000
²
$GMXVWPHQW3URFHGXUH
Figure 12. Probe Tip Adjustment Locations
12. Remove the BNC cable from the output of the Function
Generator and the calibration fixture. Leave the BNC cable
from the SYNC output to the oscilloscope external trigger
input in place.
D. Adjust OFFSET (R330)
1. Reconnect the BNC cable to both ends of the ProBus
extension cable. The oscilloscope end of the extension
cable should still be connected to channel 1.
2. Set the oscilloscope channel 1 scale factor to 10 mV/div,
and the acquisition mode to Auto. Set the BWL to 20 or
25 MHz. Adjust the OFFSET to 0.0 mV
3. Carefully move the AP034 probe tip from the Common
Mode Drive No Termination position of the AP033/AP034
Calibration Fixture to the Differential Drive No
Termination position. Press the probe into the fixture to
fully engage the pins.
4. Connect female BNC-to-dual male banana plug adapters to
each end of the BNC cable. Plug one end into the output of
the DC voltage source, making sure the plug corresponding
to the BNC shield (marked “Ground”) is connected to the
voltage source –Output (or common for dual supplies)
connector, and the other pin on the +Output connector.
C17A
Attenuator Comp
C18A
Attenuator CMRR
R7
LF CMRR
Summary of Contents for AP034
Page 1: ... QVWUXFWLRQ 0DQXDO H UR 3 FWLYH LIIHUHQWLDO 3UREH REVISION D JANUARY 2000 ...
Page 6: ...YL ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 10: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 26: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 68: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 72: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH Figure 13 Typical CMRR Graphs ...
Page 76: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...