C. Check Low Frequency CMRR
Note
Because greater amplitudes are required
to measure the higher CMRR
specifications at low frequencies, the
Function Generator will be used in place of
the high-frequency sine wave generator
for the low frequency CMRR test.
1. Carefully move the AP034 probe head from the Common
Mode Drive 50 ohm Termination connector of the
AP033/AP034 Calibration Fixture to the Common Mode
Drive No Termination connector. Make that sure the probe
is fully engaged in the fixture.
2. Set the oscilloscope to display channel 2, the channel 2 input
coupling to DC1 M
Ω
, the channel 2 vertical scale to
5 Volt/div, the horizontal scale to 5 ms/div, and the trigger
source to channel 2. Set BW limiting on channels 1 and 2 to
25 MHz.
3. Connect the BNC Tee adapter to the output connector of the
Function Generator. Connect a BNC cable from one end of
the BNC Tee adapter to channel 2 of the oscilloscope.
Connect a second BNC cable from the remaining end of the
BNC Tee adapter to the Common Mode Drive No
Termination input connector of the AP033/AP034
Calibration Fixture.
Caution
Make sure that the Common Mode Drive
No Termination connection is used.
Prolonged application of the power levels
used in the low frequency common mode
test may damage the termination
resistance in either 50 ohm Termination
input of the AP033/AP034 calibration
fixture.
Summary of Contents for AP034
Page 1: ... QVWUXFWLRQ 0DQXDO H UR 3 FWLYH LIIHUHQWLDO 3UREH REVISION D JANUARY 2000 ...
Page 6: ...YL ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 10: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 26: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 68: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 72: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH Figure 13 Typical CMRR Graphs ...
Page 76: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...