![LeCroy AP034 Instruction Manual Download Page 34](http://html1.mh-extra.com/html/lecroy/ap034/ap034_instruction-manual_1866287034.webp)
²
ISSUED: January 2000
AP034-OM-E Rev D
$3$FWLYH3UREH
MATCHING PROCEDURE FOR
÷
10 PLUG-ON EXTERNAL ATTENUATOR
The
÷
10 and
÷
20 Plug-on attenuators provided as standard
accessories with the AP034 are calibrated to match the specific
probe they are shipped with. Individual probes will have small
variations in parasitic capacitance within the input circuits. To
obtain maximum common mode rejection performance, the
attenuators are calibrated to match a specific probe during the
manufacturing process. In order to preserve the maximum
Common Mode Rejection, do not interchange external
attenuators between probes.
The Plug-on AC coupling adapter is not matched to a specific
probe and, therefore, does not need to be matched when
interchanged.
If the
÷
10 or
÷
20 Plug-on attenuators become accidentally mixed
between probes, you can use the procedure listed below to
restore the compensation match. This adjustment does not affect
any of the parameters in the warranted specifications. Therefore,
the required test equipment does not need to be calibrated.
Note
The AP033 and AP034 Active Differential Probes have
different input capacitance. The
÷
10 Plug-on attenuator
supplied with model AP033 cannot be properly adjusted
for use with model AP034. Make sure that the attenuator
is marked “AP034” before attempting this procedure.
EQUIPMENT REQUIRED
Test Oscilloscope
The oscilloscope must support ProBus. Otherwise use a non-
ProBus oscilloscope and ADPPS power supply.
Signal Source
Low frequency square wave: frequency from 50 Hz to 5 kHz,
amplitude from 1 V to 10 V. The output waveform must have a
square corner and flat top with minimum overshoot suitable for
adjusting compensation. The generator should have trigger
output, or use a BNC Tee connector and separate BNC cable
Summary of Contents for AP034
Page 1: ... QVWUXFWLRQ 0DQXDO H UR 3 FWLYH LIIHUHQWLDO 3UREH REVISION D JANUARY 2000 ...
Page 6: ...YL ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 10: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 26: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 68: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...
Page 72: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH Figure 13 Typical CMRR Graphs ...
Page 76: ... ISSUED January 2000 AP034 OM E Rev D 3 FWLYH 3UREH BLANK PAGE ...